GrayWolf气体检测仪临时性过滤膜湿度测试实验总结
将型号为TG502的格雷沃夫气体检测仪探头放置到湿度箱中,分别采集不加装过滤膜和加装过滤膜时,在湿度箱中的数据,间隔时间为10S。得出如下所示数据:
未装过滤膜部分数据:
Date Time | TVOC ppb | Temperature °C | Relative Humidity %RH |
27-Nov-14 02:01:04 AM | 166 | 16.8 | 36.8 |
27-Nov-14 02:01:14 AM | 167 | 16.8 | 39.7 |
27-Nov-14 02:01:24 AM | 169 | 16.9 | 44.3 |
27-Nov-14 02:01:34 AM | 172 | 16.9 | 47.8 |
27-Nov-14 02:03:34 AM | 190 | 17.4 | 74.8 |
27-Nov-14 02:03:44 AM | 193 | 17.4 | 76.3 |
27-Nov-14 02:04:44 AM | 207 | 17.7 | 83.0 |
27-Nov-14 02:04:54 AM | 207 | 17.7 | 83.9 |
27-Nov-14 02:05:04 AM | 209 | 17.8 | 84.8 |
27-Nov-14 02:07:14 AM | 237 | 18.2 | 93.1 |
27-Nov-14 02:07:24 AM | 239 | 18.3 | 93.6 |
27-Nov-14 02:07:34 AM | 238 | 18.3 | 94.0 |
27-Nov-14 02:09:54 AM | 247 | 18.7 | 97.4 |
27-Nov-14 02:10:04 AM | 248 | 18.8 | 97.5 |
装过滤膜部分数据:
Date Time | TVOC ppb | Temperature °C | Relative Humidity %RH |
27-Nov-14 02:14:26 AM | 150 | 18.6 | 32.7 |
27-Nov-14 02:14:36 AM | 150 | 18.8 | 36.7 |
27-Nov-14 02:14:46 AM | 155 | 18.9 | 51.0 |
27-Nov-14 02:18:36 AM | 211 | 19.5 | 88.2 |
27-Nov-14 02:18:46 AM | 212 | 19.6 | 88.5 |
27-Nov-14 02:20:56 AM | 215 | 19.8 | 90.4 |
27-Nov-14 02:22:36 AM | 221 | 19.9 | 91.2 |
27-Nov-14 02:30:16 AM | 213 | 20.3 | 92.7 |
27-Nov-14 02:30:26 AM | 213 | 20.3 | 92.7 |
27-Nov-14 02:30:36 AM | 212 | 20.3 | 92.7 |
27-Nov-14 02:30:46 AM | 211 | 20.3 | 92.7 |
27-Nov-14 02:30:56 AM | 211 | 20.3 | 92.7 |
27-Nov-14 02:31:06 AM | 212 | 20.3 | 92.7 |
1、不装过滤膜采集得到数据中,湿度最小值为36.8%,湿度最大值为97.5%
2、装过滤膜采集得到的数据中,湿度最小值为32.7%,湿度最大值为92.7%
3、不装过滤膜采集数据总时长为10分钟
4、装过滤膜采集数据总时长为17分钟
5、在相同环境中,装过滤膜时采集的最小湿度比不装过滤膜采集的最小湿度值低4.1%;
装过滤膜采集的最大湿度比不装过滤膜采集的最大湿度值低4.8%
6、在相同环境中,装过滤膜采集数据的时长要比不装过滤膜采集数据的时长多出7分钟,即装过滤膜采集数据的时间要比不装过滤膜采集数据的时间长。
综合以上内容及数据,可以得出:GrayWolf气体检测仪探头临时性过滤膜会对湿度产生影响。